Design of Respiratory Sensor System for polysomnography using Thermocouple

Polysomnography를 위한 열전대 호흡센서 시스템의 설계

  • 우용규 (서울대학교 대학원 협동과정 의용생체공학전공) ;
  • 정도언 (서울대학교 의과대학 정신과학교실) ;
  • 박광석 (서울대학교 의과대학 의공학교실)
  • Published : 2000.06.01

Abstract

Changes in breathing pattern and apnea both can be !he result of sleep disorders. The focus of this paper is to develop methodologies to monitor the breathing pattern and to detect apnea. An accurate recording of the respiratory phase can be carried out with different methods. One of these methods is the use of a thermocouple, which reacts to the variation in air temperature, placed in the nose and mouth of the patient. The K-type thermocouple was used because it has high reliability, thermo-stability, and good corrosion resistance. And also, it has a considerable long time constant that gives a low cut-off frequency, well below the respiratory frequency and thereby causing a large phase difference. The result showed that timing of respiration was accurately obtained with the AD595, amplifier for K-type thermocouple.

Keywords