원자 현미경 장비의 바닥 진동(정상 상태) 허용 기준 결정

Determination of the Allowable Vibration Level of the Atomic Force Microscope Equipment

  • 발행 : 2000.11.16

초록

Currently, Atomic Force Microscope(AFM) has been widely used to measure the surface topography of a sample by detecting interaction force between atoms on the sample and extremely sharp probe tip. The vertical resolution of AFM is mainly determined by external vibration noise. The resolution of AFM shows different values for the different environment, thus it is necessary to determine relationship between the criteria and the resolution of AFM regardless of environment. In this paper, we discuss the allowable level of floor vibration for AFM equipment at given resolution. The vibration criteria can be used as reference data to design mechanical structure and to analyze the structural dynamics of AFM equipment.

키워드