Cancellation of Motion Artifact in MRI Image Plane
MRI 촬상단면내의 체동 아티팩트의 제거
Abstract
In this study, a new algorithm for canceling MRI artifacts through the translational motion of image plane is presented. Bloating is often makes problems in a clinical diagnosis. Assuming that the head moves up and down due to breathing, rigid translational motions in only y(phase encoding axis) direction is treated. First, we notice that the x directional motion corresponds to a shift of the x directional spectrum of the MRI signal, and the non zero area of the spectrum just corresponds to x axis projected area of the density function. So the motion is estimated by tracing the edges of the spectrum, and the x directional motion is canceled by shifting the spectrum in inverse direction. Next, the y directional motion is canceled using a new constraint, with which the motion component and the true image component can be separated. Finally, the effectiveness of this algorithm is shown by using a phantom with simulated motions.
Keywords