Design of Z-directional gradient coil to improve gradient linearity for the nuclear magnetic resonace imaging(NMRI)

경사자장의 선형성 향상을 위한 핵자기공명 영상용 Z-방향 경사자기장 코일 설계

  • Published : 1998.07.20

Abstract

Gradient coil offers the spatial informations of sample or patient in Nuclear Magnetic Resonance Imaging(NMRI) and its gradient field linearity over the field of view(FOV) has many influence on the MR imaging. Accurate and good quality MR imaging can be acquired by the high gradient field linearity over the FOV. So it is an important part to design of gradient coil with good linearity in the wide imaging range. Usually, Z-directional gradient field is generated by using the Helmholtz type coil which is consisted of one-pair loop with anti-current path. It gets less about 40% linearity of the diameter spherical volume(DSV). In this study, we calculated optimized geometrical parameters of two-pair loop system to cancel odd terms up to $B_7$ included effectively. we also analyzed and compared the gradient field distribution and linearity of the common Helmholtz coil with them of the two-pair loop system.

Keywords