Development of Simulator for Evaluation Balance by Using Vestibuloocular Reflex

전정안구반사를 이용한 평형감각 평가용 시뮬레이터의 개발

  • 임승관 (원광대학교 전자공학과) ;
  • 김규겸 (원광대학교 전자공학과) ;
  • 정호춘 (원광대학교 전자공학과) ;
  • 진달복 (원광대학교 전자공학과) ;
  • 김민선 (원광대학교 의과대학 생리학교실) ;
  • 박병림 (원광대학교 의과대학 생리학교실)
  • Published : 1997.11.01

Abstract

The purpose of this study was to develop a sinusoidal rotatory chair systim for evaluating the vestibular function in patients suffered from vertigo with vestibualr disorders. The sinusoidal rotatory chair system is composed of a rotatory chair systim and softwares. Maximum velocity of the rotatiry chair was upto 60 degree per second and frequency range was 0.01 to 0.64 Hz. To evaluate the vestibular function in vertigo patients, vestibuloocular reflex was measured by sinusoidal rotation of the whole body about vertical axis in the darkness, and optokinetic nystagmus, visual vestibuloocular reflex , and visual supression test were also performed. Eye movement was measured by means of a electronystgmograph amplifier through Ag-AgCI surface electrodes. Gain, phase, and symmetry were obtained from analysis of the eye movement ineuced by vestibular or visual stimulation. In healthy adults, sinusoidal rotation of the shole body produced nystagmus, of which gain was directly proportional to the velocity of stimulation. The gain of visual vestibuloocular reflex or optokinetic nystagmus, and it was suppressed markedly by visual suppression test. These results suggest that the sinusoieal rotatory chair systim which was developed by this study can evaluate the vestibular function quantitatively, also this system is very useful to diagnose and to dstimate the vestibualr function during recovery from vertigo

Keywords