Infusion Pump용 Drop Sensor 개발에 관한 연구

  • Published : 1995.11.17

Abstract

When we inject drugstuffs to a patient for a long time, it is important to control proper injection amount and flow rate. Since inproper injection amount and a flow rate would cause bad a recovery a patient, the relience of sensors which detect injection amount is an important factor for whole injection systems' performance. In this research, we've compared the suitability of three sensors for injection pump monitoring system. The three types of sensors, piezo film sensor, photo transistor made up with three transmitting photodiodes and receving photodiode, and photo array, were selected for comparing. Using suggested data processing technique and photo amy sensors, we could minimize the effect of interference, disturbance, illumanation change, and sensitivity change caused by sensor's position. According to the experiments, the photo amy showed the higher reliance than any other the three types of sensors. The developed systems could be the foundation of beginning home production of infusion pump system and available for the base model of whole monitoring and control systems.

Keywords