실시간처리를 위한 태아심음 추출 알고리듬

Fetal heart rate estimation algorithm for real-time processing

  • Lee, Eung-Gu (Dept. of Electronic Engineering, Hanyang University) ;
  • Lee, Yong-Hui (Dept. of Electronic Engineering, Hanyang University) ;
  • Kim, Seon-Il (Dept. of Medical Information and Management University) ;
  • Lee, Du-Su (Dept. of Electronic Engineering, Hanyang University)
  • 발행 : 1994.05.13

초록

Despite the simplicity of processing, a conventional autocorrelation function(ACF) method for the precise determination of fetal heart rate (FHR) has many problems. In case of weak or noise corrupted Doppler ultrasound signal. the ACF method is very sensitive to the threshold level and data window length. It is very troublesome to extract FHR when there is a data loss. To overcome these problems, the high resolution pitch detection algorithm was adopted to estimate the FHR. This method is more accurate, robust and reliable than the ACF method. With a lot of calculation, however, it is impossible to process real time FHR estimation. This paper is presented a new FHR estimation algorithm for real time processing.

키워드