Under-Relaxed Image Restorative Technique for $Na^{23}$ MRI

  • Ro, D.W. (Division of Medical Instrumentations Korea Academy of Industrial Technology) ;
  • Ahn, C.B. (Division of Medical Instrumentations Korea Academy of Industrial Technology)
  • Published : 1992.05.15

Abstract

To improve signal-to-noise ratio in sodium image, short echo time (2-3 ms) and long data acquisition (10-20 ms) protocols are used. Sodium in biological specimens demonstrates a bi-exponential decay of transverse magnetization and the fast decaying component of the sodium signal results in the reconstruction of images which are blurred significantly. The spatially-dependent nature of the blurs are due mainly to the presence of short local transverse relaxation values (0.7-3 ms) of sodium in tissue. We present an algorithm that corrects for object-dependent blurs due to fast-decaying T2 and improves the computational behavior of the algorithm by incorporating a relaxation parameter into the iterative process.