DOI QR코드

DOI QR Code

Calibration Technique for Freehand 3-D Ultrasound System

3차원 초음파 시스템의 캘리브레이션 방법

  • Hwang, Myun Joong (School of Mechanical & Automotive Engineering, Halla University)
  • 황면중 (한라대학교 기계자동차공학부)
  • Received : 2013.08.30
  • Accepted : 2013.12.05
  • Published : 2013.12.31

Abstract

This paper proposes a calibration method for a freehand 3-D ultrasound system in medical robotic research. The calibration block with six wires was designed to set the fixed target points. The positions of the ultrasound probe and calibration block were measured using an optical tracker. The relationship between the position of the ultrasound probe and the pixel coordinates in the images was derived using a calibration process. The scaling matrix was also calculated. The experimental results showed that the proposed method could find solutions using a simple least square method from one or multiple ultrasound images.

Keywords

Calibration;Image-guidance robot;Ultrasound

References

  1. K. Cleary, A. Melzer, V. Watson, G. Kronreif, and D. Stoianovici, "Interventional robotic systems: Applications and technology state of-the-art," Minimally Invasive Therapy, vol. 15, no. 2, pp. 101-113, 2006. DOI: http://dx.doi.org/10.1080/13645700600674179 https://doi.org/10.1080/13645700600674179
  2. Y. H. Huang, T. H. Wu, M. H. Lin, C. C. Yang, W. Y. Guo, Z. J. Wang, C. L. Chen, and J. S. Lee, "An automated robot arm system for small animal tissue biopsy under dual-image modality," Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 569, no. 2, pp. 230-234, December 2006. DOI: http://dx.doi.org/10.1016/j.nima.2006.08.072 https://doi.org/10.1016/j.nima.2006.08.072
  3. R. C. Susil, J. H. Anderson, and R. H. Taylor, "A Single Image Registration Method for CT Guided Interventions," Lecture Notes in Computer Science, vol. 1697, pp. 798-808, 1999. DOI: http://dx.doi.org/10.1007/10704282_87 https://doi.org/10.1007/10704282_87
  4. S. P. Dimaio, E. Samset, G. Fischer, I. Iordachita, G. Fichtinger, F. Jolesz, and C. M. Tempany, "Dynamic MRI Scan Plane Control for Passive Tracking of Instruments and Devices," Lecture Notes in Computer Science, vol. 4792, pp. 50-58, 2007. DOI: http://dx.doi.org/10.1007/978-3-540-75759-7_7 https://doi.org/10.1007/978-3-540-75759-7_7
  5. P. Abolmaesumi, S. Salcudean, W. Zhu, M. Sirouspour, and S. DiMaio, "Image guided control of a robot for medical ultrasound," IEEE Transactions on Robotics and Automation, vol. 18, no. 1, pp. 11-23, February 2002. DOI: http://dx.doi.org/10.1109/70.988970 https://doi.org/10.1109/70.988970
  6. A. C. Waspe, J. C. Lacefield, and A. Fenster, "Registration of three dimensional high-frequency ultrasound images to a robotic needle positioning system for preclinical research," Proc. of the 4th IEEE International Symposium on Biomedical Imaging: From Nano to Macro (ISBI), Arlington, USA, April 12-15, pp. 1132-1135, 2007. DOI: http://dx.doi.org/10.1109/ISBI.2007.357056 https://doi.org/10.1109/ISBI.2007.357056
  7. L. Mercier, T. Lango, F. Lindseth, and L. D. Collins, "A Review of Calibration Techniques for Freehand 3-D Ultrasound Systems," Ultrasound in Medicine & Biology, vol. 31, no. 2, pp. 143-165, 2005. DOI: http://dx.doi.org/10.1016/j.ultrasmedbio.2004.11.001 https://doi.org/10.1016/j.ultrasmedbio.2004.11.001
  8. NDI, Polaris Vicra Optical Tracker System, Northern Digital Inc., Waterloo, Ontario, Canada, 2009. Available From: http://www.ndigital.com/medical/polarisfamily -techspecs.php