DOI QR코드

DOI QR Code

Application of the Detection of External Contamination on Radiation Workers for Bed Type Whole Body Counting Using Monte Carlo Method

몬테카를로 방법을 적용한 bed type 전신계측기의 방사선작업종사자 외부오염 검출 응용

  • Kim, Jeong-In (Radiation Health Research Institute of Korea Hydro & Nuclear Power Co., Ltd.) ;
  • Lee, Byoung-Il (Radiation Health Research Institute of Korea Hydro & Nuclear Power Co., Ltd.)
  • 김정인 (한국수력원자력(주) 방사선보건연구원) ;
  • 이병일 (한국수력원자력(주) 방사선보건연구원)
  • Received : 2013.12.09
  • Accepted : 2013.12.18
  • Published : 2013.12.30

Abstract

Monte Carlo method was applied to discriminate the external contamination on radiation workers in nuclear power plants for internal dose assessment generally used with a bed type scanning detector whole body counter. Korean voxel model with internal contamination was used to estimate the detection patterns of whole body scanning. Also, the BOMAB model with various external contamination was assumed to compare with detection of radionuclides inside the human body. From the comparison of detection efficiency between front and back side up, external contamination was easily distinguished.

References

  1. International Commission on Radiation Protection. Individual monitoring for internal exposure of workers replacement of ICRP publication 54. ICRP Publication 78. 1997.
  2. American National Standards Institute, Inc. An American national standard - performance criteria for radiobioassay. ANSI/HPS N13.30. 1996.
  3. KIM JI et al. Efficiency calibration of bed type whole body counter using Monte Carlo simulations and application to intake estimation of $^{131}I$. Progress in Nuclear science and Technology. 2011;1;549-551. https://doi.org/10.15669/pnst.1.549
  4. Lee C et al. Development of the two Korean adult tomographic computational phantoms for organ dosimetry. Medical Physics. 2006;33(2);380-390. https://doi.org/10.1118/1.2161405
  5. Kramer GH, Burns LC, Noel L. The BRMD BOMAB phantom family. Health Phys. 1991;61(6);895-902. https://doi.org/10.1097/00004032-199112000-00025
  6. Los Alamos National Laboratory. MCNPX$^{TM}$ user's manual version 2.6.0. LA-CP-07-1473. 2008.
  7. Kramer GH et al. Monte Carlo simulation of a scanning detector whole body counter and the effect of BOMAB phantom Size on the calibration. Health Phys. 2002;83(4);526-533. https://doi.org/10.1097/00004032-200210000-00011