Abstract

This study performed and evaluated the performance test in 40 general X-ray units among diagnostic X-ray units, which were being used in hospitals of gyeongsangnam-do gimhae-si through X-ray tube variable limiting device ability test, the light field and X-ray field alignment test and collimation and beam alignment test of diagnostic X-ray unit. The results are as followings: In a variable beam limiting device ability test, the result of maximum X-ray field test showed that 4(10%) of were incongruent while the result of minimum X-ray field test represented that 5(12.5%) of were incongruent. The result of the light field and X-ray field alignment test showed 23(57.5%) of were within 2% of maximum permissible level and the other 17(42.5%) units were misalignment. The result of beam alignment test represented that 11(27.5%) coincided and another 11(27.5%) within $0.5^{\circ}$ respectively, 10(25%) were $0.6^{\circ}-1.5^{\circ}$ intervals, 7(17.5%) were $1.6^{\circ}-3^{\circ}$ and 1(2.5%) were more than $3^{\circ}$.

Keywords

Variable Beam Limiting Device;Light Field;X-ray Field;Beam Alignment

References

  1. 한국공업규격 KS A 4732, 진단용 X선 가변조리개, 1982.
  2. 김정민, 김성철, 고신관, 방사선 기기 & 관리실험,대학서림출판사, p.189, 2000.
  3. 강세식, 고신관, 권달관, 권덕문, 김정민, 김흥태, 나수경, 윤철호, 이덕규, 이성길, 이종석, 홍찬규, 방사선기기 정도관리 및 실험, 정문각출판사, p.51, 2000.
  4. http://www.gammex.com
  5. 보건복지부령 제349호, 진단용 방사선 발생장치의 안전관리에 관한 규칙(제4조 1항), 2006.
  6. AAPM report, Basic Quality in Diagnostic Radiology, No.4, 1978.
  7. NCRP report, Quality Assurance for Diagnostic Imaging Equipment, No.99, 1988.
  8. Collimator and Beam Alignment Test Tool User's Guide Hand Book, Gammex, pp.2-6, 2007.
  9. 김영일, 진료영상기기 Q.C, 대학서림출판사, pp.205-213, 1996.
  10. 김성수, 허준, 이선숙, 이인자, 김성철, "진단용 X선장치의 콜리메타의 실태조사", 방사선기술학회지, 제17권, 제1호, p.70, 1994.
  11. 강세식, 권덕문, 김영근, 김창규, 이성길, 장영일, 홍찬규, 진단용 X선장치 정도관리실험, 청구출판사, p.239, 2006.

Cited by

  1. Nationwise Survey of the X-ray Beam Collimator Utilization in General Diagnostic Radiograph vol.24, pp.2, 2013, https://doi.org/10.14316/pmp.2013.24.2.119