Domestic Intercomparison Study for the Performance of Personnel Dosimeters

개인선량계 성능의 국내 상호비교

  • Published : 1996.09.30

Abstract

The Korea Atomic Energy Research Institute(KAERI) conducted a intercomparison study for personnel dosimetry services in Korea to enhance the accuracy and precision of the dosimetry system. Nine types of dosimeters(6 TLD, 3 film badge) from 7 institutions took part in this intercomparison study. Each participant submitted 30 dosimeters including transit control for irradiations. Both TLDs and film badges were irradiated with Cs-137 gamma, Sr/Y-90 beta and 4 X-ray beams in ISO wide series. Four dosimeters were irradiated on phantom with same dose equivalent for each field category. The delivered dose equivalent was in the range of $0.1{\sim}10mSv$. The participants assessed the results of their dosimeter readings in terms of the ICRU operational quantities for personal monitoring, Hp(10) and Hp(0.07). Most participants except 1 dosimeter estimated the delivered dose equivalent with biases less than ${\pm}25%$ for Cs-137 and Sr/Y-90. But for X-rays, the biases exceeded ${\pm}35%$ in some cases bacause the dose evaluation algorithm was based on the ANSI N13.11 X-ray fields which are different from those given by ISO.